Tips for weighing sensor replacement
- Categories:Technical knowledge
- Time of issue:2018-01-24 16:10
(Summary description)Step 1 I. Disconnect the power supply ii. take off the table top iii. loosen and remove the screws connecting the sensor to the upper scale body iv. remove the mounting plate V. Turn the scale over and remove the attachment screws at the back Remove the sensor vii. install a new sensor of the same type in the reverse order Step 2 viii. Adjust the overload protection (check that the clearance at the six overload points is up to standard using a plug gauge) ix. Pass the restraint through the column from the bottom wire section x. Tuck the excess harness into the column from the bottom (pay attention to corner handling) XI. Connect the sensor harness xii. Check the corner difference xiii. Re-calibrate (see Instrument Commissioning Training)
Tips for weighing sensor replacement
(Summary description)Step 1
I. Disconnect the power supply
ii. take off the table top
iii. loosen and remove the screws connecting the sensor to the upper scale body
iv. remove the mounting plate
V. Turn the scale over and remove the attachment screws at the back
Remove the sensor
vii. install a new sensor of the same type in the reverse order
Step 2
viii. Adjust the overload protection (check that the clearance at the six overload points is up to standard using a plug gauge)
ix. Pass the restraint through the column from the bottom wire section
x. Tuck the excess harness into the column from the bottom (pay attention to corner handling)
XI. Connect the sensor harness
xii. Check the corner difference
xiii. Re-calibrate (see Instrument Commissioning Training)
- Categories:Technical knowledge
- Time of issue:2018-01-24 16:10
- Views:
Step 1
I. Disconnect the power supply
ii. take off the table top
iii. loosen and remove the screws connecting the sensor to the upper scale body
iv. remove the mounting plate
V. Turn the scale over and remove the attachment screws at the back
Remove the sensor
vii. install a new sensor of the same type in the reverse order
Step 2
viii. Adjust the overload protection (check that the clearance at the six overload points is up to standard using a plug gauge)
ix. Pass the restraint through the column from the bottom wire section
x. Tuck the excess harness into the column from the bottom (pay attention to corner handling)
XI. Connect the sensor harness
xii. Check the corner difference
xiii. Re-calibrate (see Instrument Commissioning Training)